Oral-Fluid Drug Testing System 5000 (IVD)
Request a QuoteThe Oral-Fluid Drug Testing System 5000 (IVD) is a fast, accurate means of testing oral fluid samples for drugs of abuse, including amphetamines, designer amphetamines, opiates, cocaine and metabolites, benzodiazepines, cannabinoids and methadone, for law enforcement use only.
Key features
- Tests oral fluid samples for amphetamines, designer amphetamines, opiates, cocaine and metabolites, benzodiazepines, cannabinoids and methadone.
- Easy data management through the dedicated diagnostics software.
- Test kits available for 8, 7, 6 or 5 substances (cocaine, opiates, benzodiazepines, cannabis, amphetamine, methamphetamines, methadone and ketamine, depending on the kit), supplied in packs of 20.
- Training test kits available (3 positive and 3 negative), along with a sample-collection kit pack of 20 pieces.
- Optional accessories: compact keyboard (German, English or French layout), barcode reader, mobile printer, carrying case and hard shell carrying case.
- Analyser supplied with power supply and 12 V car cable (IVD version).
- Simultaneous qualitative detection of substances or substance classes in human saliva for diagnostic purposes (in vitro diagnostics) and forensic use; the detectable substances are defined by the test kit used.
- Qualitative screening method that detects target substances or their metabolites above a specified threshold concentration (cut-off); a confirmed result requires a more specific method, preferably gas chromatography/mass spectrometry (GC/MS).
- Professional evaluation of results must consider the broader clinical context of the individual tested, particularly for a preliminary positive result.
Technical data
- Operating voltage: 12 V direct voltage (via supplied power pack)
- Power input: normally 3 A
- Dimensions (W x H x D): 200 x 260 x 250 mm
- Weight: 4.5 kg
- Operating temperature: 4 °C to 40 °C
- Storage / transportation temperature: -20 °C to +60 °C
- Relative humidity: 5 to 95 % (non-condensing)
- Ports: optical for printer, PS/2, USB Slave
- Time of measurement: depending on the test kit used (typically < 9 minutes); optional analysis in < 5 minutes by increasing the THC cut-off to 25 ng/ml (after updating the device software)
- Storage capacity: 500 records or measurements
- Licences / standards: Licensed as a medical product within the EU in accordance with Directive 98/79/EC on in vitro diagnostic medical devices (IVD). Outside the European Union for non-medical use (Standard) or for forensic use (UKCA-conform).
Supplied, installed, trained and serviced across India by Dawell Lifescience, with calibration and AMC support.
₹ 11,99,000.00
₹ 11,99,000.00
₹ 11,41,904.76