DrugCheck 3000 (disposable)

device photo

Dräger DrugCheck 3000 — field card

Use it for: Instrument-free oral-fluid screening where no analyser is available.

Steps

  1. Nothing in the mouth for 10 minutes; 5–40 °C; gloves on.
  2. Person moves the collector between gums and cheeks ~15 s until the red indicator loses colour (max 2 min).
  3. Push the collector fully into the cassette to break the ampule; shake well.
  4. Read in good light after the time on the pack.
  5. Control line present + drug line present = negative; no drug line = positive ("not negative"); no control line = invalid.

It cannot

  • Give a quantity or impairment.
  • Be read by line strength — only presence or absence counts.
  • Replace lab confirmation.

Law link: Same as other oral-fluid screens: MV Act s.185(b) / NDPS s.27 need lab confirmation and recorded observations.

Dawell Lifescience Pvt. Ltd. · Training & support: narcotics@dawelllifescience.com · +91 91563 19641 · Field results are presumptive screening; identity and quantity are decided by the FSL report. After training, Dawell issues a Training Completion Certificate.