DrugCheck 3000 (disposable)

Dräger DrugCheck 3000 — field card
Use it for: Instrument-free oral-fluid screening where no analyser is available.
Steps
- Nothing in the mouth for 10 minutes; 5–40 °C; gloves on.
- Person moves the collector between gums and cheeks ~15 s until the red indicator loses colour (max 2 min).
- Push the collector fully into the cassette to break the ampule; shake well.
- Read in good light after the time on the pack.
- Control line present + drug line present = negative; no drug line = positive ("not negative"); no control line = invalid.
It cannot
- Give a quantity or impairment.
- Be read by line strength — only presence or absence counts.
- Replace lab confirmation.
Law link: Same as other oral-fluid screens: MV Act s.185(b) / NDPS s.27 need lab confirmation and recorded observations.