TruNarc Tau handheld narcotics analyser
Thermo Scientific TruNarc Tau â field card
Use it for: Fast, single-operator screening of suspected narcotics, precursors and cutting agents â through clear plastic bags and thin containers.
Steps
- Power on and enter the laser activation code (issued after training).
- Self-check standard open (down) â closed only for the self-check.
- Press the nose cone into the bag so the laser spot sits in the substance (at least a rice-grain amount).
- Select Scan; keep still until the laser light goes off.
- Read: Alarm (controlled substance), Clear (cutting agent), Precursor, Inconclusive.
- Type H kit for low-concentration heroin/fentanyl; note "fentanyl and/or methamphetamine" result needs lab.
It cannot
- Scan trace residue films.
- Confirm identity for court.
- Clear an "acetaminophen" result â narcotics may be below detection.
Law link: Weigh and seal before testing; record the result in the panchnama and Form-1 (Rule 3). Never point the laser at eyes.
Dawell Lifescience Pvt. Ltd. ¡ Training & support: narcotics@dawelllifescience.com ¡ +91 91563 19641 ¡ Field results are presumptive screening; identity and quantity are decided by the FSL report. After training, Dawell issues a Training Completion Certificate.