TruNarc Tau handheld narcotics analyser

Thermo Scientific TruNarc Tau — field card

Use it for: Fast, single-operator screening of suspected narcotics, precursors and cutting agents — through clear plastic bags and thin containers.

Steps

  1. Power on and enter the laser activation code (issued after training).
  2. Self-check standard open (down) — closed only for the self-check.
  3. Press the nose cone into the bag so the laser spot sits in the substance (at least a rice-grain amount).
  4. Select Scan; keep still until the laser light goes off.
  5. Read: Alarm (controlled substance), Clear (cutting agent), Precursor, Inconclusive.
  6. Type H kit for low-concentration heroin/fentanyl; note "fentanyl and/or methamphetamine" result needs lab.

It cannot

  • Scan trace residue films.
  • Confirm identity for court.
  • Clear an "acetaminophen" result — narcotics may be below detection.

Law link: Weigh and seal before testing; record the result in the panchnama and Form-1 (Rule 3). Never point the laser at eyes.

Dawell Lifescience Pvt. Ltd. ¡ Training & support: narcotics@dawelllifescience.com ¡ +91 91563 19641 ¡ Field results are presumptive screening; identity and quantity are decided by the FSL report. After training, Dawell issues a Training Completion Certificate.