DrugTest 5000 + SSK 5000

Dräger DrugTest 5000 + SSK 5000 — field card
Use it for: Instrument oral-fluid screening of a person; SSK 5000 swab screening of surfaces (dashboards, tables, packaging).
Steps
- Person: nothing in the mouth for 10 minutes; cassette at room temperature; open the pouch only now.
- Person moves the sampler cheek to cheek (no chewing or sucking) until the indicator turns blue (max 4 min).
- Put the safety cap back, insert into the analyser, wait for the plain-text result; print.
- Surface (SSK 5000): photograph the area, swab a defined area, return swab to the buffer vial, dose the cassette as the IFU states.
- Keep the SSK vial sealed and labelled — it is the confirmation sample.
It cannot
- Prove consumption from a surface swab — it shows contamination only.
- Replace NDPS Rule 11 samples (24 g / 5 g).
- Confirm a positive without a laboratory.
Law link: Record lot, expiry, result and video; send the confirmation sample with chain of custody.
Dawell Lifescience Pvt. Ltd. · Training & support: narcotics@dawelllifescience.com · +91 91563 19641 · Field results are presumptive screening; identity and quantity are decided by the FSL report. After training, Dawell issues a Training Completion Certificate.